PERBEDAAN JUMLAH TROMBOSIT CARA MANUAL PADA PEMBERIAN ANTIKOAGULAN EDTA KONVENSIONAL(PIPET MIKRO) DENGAN EDTA VACUTAINER

作者: Charles_King Wijaya

DOI:

关键词: Drop volumeVacutainerBlood samplingChemistrySignificant differenceSurgeryChromatography

摘要: Background : Pasteur pipets as generally used on conventional EDTA addition certainly result in significant difference of platelet count than vacutainer EDTA. This is caused by the overdose that spuriously increases count. One way reducing possibility making mistake with still using use which drop volume correctly needed. Micropipet a solution. Objectives To prove there was no (micropipet) and addition. Method study an observational analytic cross sectional approach. Thirty seven samples from both kinds inclusion criterias were unstatic blood sampling, rightly 3ml for EDTA, partly coagulated blood, less 2 hours storage room temperature. Computer programme processing data. Hypothesis test paired-samples t test. Result The mean lower 269594,59±29489,367/mm3 273918,92± 29607,036/mm3. Paired-samples resulted p= 0.000. Conclusion There between addition. Keyword micropipet, vacutainer,