An implantable CMOS circuit interface for multiplexed microelectrode recording arrays

作者: J. Ji , K.D. Wise

DOI: 10.1109/4.121568

关键词: Communication channelAmplifierElectrical engineeringEquivalent inputMaterials scienceMicroelectrodeMultiplexingNoise (electronics)CMOSElectrode

摘要: A second-generation multichannel probe designed for measuring single-unit activity in neural structures is described. The includes CMOS circuitry electronically positioning the recording sites with respect to active neurons and amplifying multiplexing recorded signals. selects eight from among 32 on shank using a static input channel selector. signals selected channels are then amplified multiplexed outside world. offers typical AC gain of 300 (15 Hz 7 kHz), DC 0.3, an equivalent noise 15 mu V rms. Operating single 5-V supply, dissipates 2.5 mW power implements selection, self-testing, data output, initialization three external leads. realized 12 masks high-yield single-sided dissolved wafer process 3- m feature size pitch electrode shanks. >

参考文章(8)
M.S.J. Steyaert, W.M.C. Sansen, A micropower low-noise monolithic instrumentation amplifier for medical purposes IEEE Journal of Solid-state Circuits. ,vol. 22, pp. 1163- 1168 ,(1987) , 10.1109/JSSC.1987.1052869
K.L. Drake, K.D. Wise, J. Farraye, D.J. Anderson, S.L. BeMent, Performance of planar multisite microprobes in recording extracellular single-unit intracortical activity IEEE Transactions on Biomedical Engineering. ,vol. 35, pp. 719- 732 ,(1988) , 10.1109/10.7273
J.-C. Bertails, Low-frequency noise considerations for MOS amplifiers design IEEE Journal of Solid-State Circuits. ,vol. 14, pp. 773- 776 ,(1979) , 10.1109/JSSC.1979.1051262
K. Najafi, K.D. Wise, T. Mochizuki, A high-yield IC-compatible multichannel recording array IEEE Transactions on Electron Devices. ,vol. 32, pp. 1206- 1211 ,(1985) , 10.1109/T-ED.1985.22102
K. Najafi, J. Ji, K.D. Wise, Scaling limitations of silicon multichannel recording probes IEEE Transactions on Biomedical Engineering. ,vol. 37, pp. 1- 11 ,(1990) , 10.1109/10.43605
M.G. Dorman, M.A. Prisbe, J.D. Meindl, A monolithic signal processor for a neurophysiological telemetry system IEEE Journal of Solid-state Circuits. ,vol. 20, pp. 1185- 1193 ,(1985) , 10.1109/JSSC.1985.1052457
J. Ji, K. Najafi, K.D. Wise, A low-noise demultiplexing system for active multichannel microelectrode arrays IEEE Transactions on Biomedical Engineering. ,vol. 38, pp. 75- 81 ,(1991) , 10.1109/10.68212
K. Najafi, K.D. Wise, An implantable multielectrode array with on-chip signal processing international solid-state circuits conference. ,vol. 21, pp. 1035- 1044 ,(1986) , 10.1109/JSSC.1986.1052646