作者: BRUNO VALTIER , BERNARD P. CHOLLEY , JEAN-PIERRE BELOT , JEAN-EMMANUEL de la COUSSAYE , JOAQUIM MATEO
DOI: 10.1164/AJRCCM.158.1.9707031
关键词: Mechanical ventilation 、 Fick principle 、 Cardiac output 、 Blood flow 、 Internal medicine 、 Critically ill 、 Intensive care 、 Artificial ventilation 、 Cardiology 、 Doppler effect 、 Medicine 、 Anesthesia
摘要: Measurement of cardiac output using thermodilution technique in mechanically ventilated patients is associated with significant morbidity. The goal the present study was to assess validity measurement transesophageal Doppler critically ill patients. Forty-six from three different intensive care units underwent 136 paired measurements (COTH) and (COTED). In addition, simultaneous suprasternal indirect calorimetry (Fick principle) were used measure 26 one center. A good correlation found between COTH COTED (r = 0.95), a small systematic underestimation (bias 0.24 L/min) Doppler. limits agreement + 2 L/min − 1.5 L/min. Variations two consecutive measures either or techniques similar direction magnitude (b...