Direct traceability of the portable QASUME irradiance scale to the primary irradiance standard of the PTB

作者: Julian Gröbner , Peter Sperfeld

DOI: 10.1088/0026-1394/42/2/008

关键词: UltravioletWavelength rangeIrradianceSpectroradiometerBlack-body radiationScale (ratio)OpticsPhysics

摘要: The portable Quality Assurance of Spectral UV Measurements in Europe (QASUME) irradiance scale has been directly compared to the primary standard (blackbody BB3200pg) Physikalisch-Technische Bundesanstalt (PTB) using travelling spectroradiometer QASUME. average blackbody spectral based on three different temperatures (2940?K, 2997?K and 3059?K) was 0.7% greater than QASUME low-power lamps. This is fully consistent with expanded uncertainty (k = 2) 4% ultraviolet (UV) wavelength range 2.6% above 400?nm scale. If these measurements are used establish scale, then its substantially reduced 2.5% 2.0% 400?nm.

参考文章(4)
P Sperfeld, J Metzdorf, S Galal Yousef, K D Stock, W M ller, Improvement and extension of the black-body-based spectral irradiance scale Metrologia. ,vol. 35, pp. 267- 271 ,(1998) , 10.1088/0026-1394/35/4/9
Julian Gröbner, Josef Schreder, Stelios Kazadzis, Alkiviadis F. Bais, Mario Blumthaler, Peter Görts, Rick Tax, Tapani Koskela, Gunther Seckmeyer, Ann R. Webb, Diana Rembges, Traveling reference spectroradiometer for routine quality assurance of spectral solar ultraviolet irradiance measurements Applied Optics. ,vol. 44, pp. 5321- 5331 ,(2005) , 10.1364/AO.44.005321