Structure of the interface between rabbit cortical bone and implants of gold, zirconium and titanium

作者: P THOMSEN , C LARSSON , L. E ERICSON , L SENNERBY , J LAUSMAA

DOI: 10.1023/A:1018579605426

关键词: MetalTransition metalMetallurgyLayer (electronics)Materials scienceBiomedical engineeringZirconiumAmorphous solidTitaniumImplantCortical bone

摘要: The role of surface properties (chemical and structural) for the interaction between biomaterials tissue is not yet understood. In present study, implants made titanium, zirconium (transition metals with oxides) gold (metallic surface) were inserted into rabbit tibia. Light microscopic (LM) morphometry showed that after 1 6 mo had less amount bone within threads a lower degree bone-implant contact than titanium implants, which did differ from each other. These quantitative differences supported by LM ultrastructural observations interface. in addition demonstrated layer non-collagenous amorphous material located implant calcified was appreciably thicker around implants. factors potentially responsible observed morphological different surfaces are discussed.

参考文章(35)
D M Brunette, The effects of implant surface topography on the behavior of cells. International Journal of Oral & Maxillofacial Implants. ,vol. 3, pp. 231- 246 ,(1988)
Arthur D. Baker, C. R. Brundle, Electron spectroscopy : theory, techniques, and applications Academic Press. ,(1977)
Sennerby L, Thomsen P, Ericson Le, A morphometric and biomechanic comparison of titanium implants inserted in rabbit cortical and cancellous bone. International Journal of Oral & Maxillofacial Implants. ,vol. 7, pp. 62- 71 ,(1992)
The Oxide handbook Published in <b>1973</b>. ,(1973) , 10.1007/978-1-4615-9597-7
David F. Williams, Titanium as a metal for implantation. Part 2: biological properties and clinical applications. Journal of Medical Engineering & Technology. ,vol. 1, pp. 266- 270 ,(1977) , 10.3109/03091907709162192
J Galante, W Rostoker, Corrosion-related failures in metallic implants: an experimental study. Clinical Orthopaedics and Related Research. ,vol. 86, pp. 237- 244 ,(1972) , 10.1097/00003086-197207000-00036
T ALBREKTSSON, H HANSSON, B IVARSSON, Interface analysis of titanium and zirconium bone implants. Biomaterials. ,vol. 6, pp. 97- 101 ,(1985) , 10.1016/0142-9612(85)90070-5