Realization of a filter radiometer-based irradiance scale with high accuracy in the region from 286 nm to 901 nm

作者: Murat Durak , Farhad Samadov

DOI: 10.1088/0026-1394/41/6/007

关键词: DetectorHalogen lampMaterials scienceRealization (systems)SiliconApertureRemote sensingRadiometerFilter (video)OpticsIrradianceGeneral Engineering

摘要: A technique is introduced for realizing spectral irradiance in the range between 286 nm and 901 nm. The scale based on a temperature-controlled filter radiometer constructed from three-element silicon trap detector, band-pass filters precision aperture. Seasoned FEL-type quartz halogen lamps are used as reference standards. In this method, direct traceability of to cryogenic gives an important advantage visible near-infrared regions. Another system procedure followed measurement distance lamp filament aperture surface radiometer. expanded uncertainty (k = 2) 2.3% 1.5% UV region, while order 0.6% above 400 nm.

参考文章(22)
M. Durak, F. Samedov, Realization of luminous flux unit lumen at National Metrology Institute of Turkey (UME Optica Applicata. ,vol. 34, pp. 265- 274 ,(2004)
J E Martin, N P Fox, P J Key, A Cryogenic Radiometer for Absolute Radiometric Measurements Metrologia. ,vol. 21, pp. 147- 155 ,(1985) , 10.1088/0026-1394/21/3/007
P Sperfeld, J Metzdorf, S Galal Yousef, K D Stock, W M ller, Improvement and extension of the black-body-based spectral irradiance scale Metrologia. ,vol. 35, pp. 267- 271 ,(1998) , 10.1088/0026-1394/35/4/9
Petri Kaerhae, Antti Lassila, Hanne Ludvigsen, Farshid Manoochehri, Heidi Fagerlund, Erkki Ikonen, Optical power and transmittance measurements and their use in detector-based realization of the luminous intensity scale Optical Engineering. ,vol. 34, pp. 2611- 2618 ,(1995) , 10.1117/12.208095
KD Stock, KH Raatz, P Sperfeld, J Metzdorf, T Kübarsepp, P Kärhä, E Ikonen, L Liedquist, Detector-stabilized FEL lamps as transfer standards in an international comparison of spectral irradiance Metrologia. ,vol. 37, pp. 441- 444 ,(2000) , 10.1088/0026-1394/37/5/21
C. L. Cromer, G. Eppeldauer, J. E. Hardis, T. C. Larason, A. C. Parr, National Institute of Standards and Technology detector-based photometric scale. Applied Optics. ,vol. 32, pp. 2936- 2948 ,(1993) , 10.1364/AO.32.002936