作者: Ali Salari , Babak Mohammadzadeh Asl
DOI: 10.1109/ICBME.2018.8703502
关键词: Adaptive beamformer 、 Minimum-variance unbiased estimator 、 Mathematics 、 Diagonal 、 Robustness (computer science) 、 Loading factor 、 Medical ultrasound 、 Algorithm 、 Anechoic chamber 、 Ultrasound
摘要: In medical ultrasound imaging, the most famous adaptive beamforming algorithm is minimum variance (MV). Diagonal loading (DL) method usually used to improve robustness of MV. Conventional DL methods have a critical problem which determining factor ignoring input data. To address this problem, in paper, using shrinkage proposed coefficient completely automated and determined by The performance evaluated simulated data Field II. point targets simulation, it has been shown that improves resolution about 87% 20%, compared delay sum (DAS) MV algorithms (with fixed of), respectively. addition, anechoic cyst contrast relative retained, comparison those beamformer, while they are improved 5% 18%, DAS ones,