作者: Alexander P. Bianchi , Trevor K.B. Joyner
DOI: 10.1016/S0021-9673(97)00098-8
关键词: Analytical chemistry 、 Gas chromatography 、 Chemistry 、 Tenax 、 Chromatography 、 Detection limit 、 Flame ionization detector 、 Ion trap 、 Thermal desorption 、 Sample preparation 、 Mass spectrometry
摘要: Abstract This work describes a method for the determination of low levels toluene-2,4-diisocyanate (TDI) in workplace air. Perkin-Elmer ATD-400 adsorbent tubes containing 200 mg Tenax-TA (80–100 mesh) were used with low-flow personal sampling pumps. Following period sampling, sealed and returned to laboratory analysis within 24 h. The TDI content was determined by thermal desorption followed capillary column gas chromatography simultaneous FID-ITD detection. analytical run time less than 30 min. Based on nominal 1-1 air sample volume, detection limit as 0.001 μg TDI.