COMPARISON OF THE PERFORMANCE OF TWO OSL READERS FOR THE PERSONAL DOSE EQUIVALENT HP(10) MEASUREMENT.

作者: Augustin Simo , Thierry Ndzana Ndah , Yolande Huguette Ebele Yigbedeck , Calvin Didier Njiki , Jean Féli Beyala Ateba

DOI: 10.1093/RPD/NCAB022

关键词: Equivalent doseSimulationRadiation protectionOptically stimulated luminescenceDosimetryComputer science

摘要: Technical performances of two optically stimulated luminescence readers including, a microStar reader and an Auto 200 both manufactured by Landauer, Inc. were studied the National Radiation Protection Agency Dosimetry Laboratory using manufacturer International Electrotechnical Commission requirements. The tools demonstrated satisfactory results as they met all showed similar performance for overall uncertainty Coefficient Variation tests. better linearity quality control tests, whereas has lower limit detection. Moreover, average shift 0.08 mSv exists when measuring doses with readers.