An ultrasound biomicroscopic dark-room provocative test.

作者: Pavlin Cj , Harasiewicz K , Foster Fs

DOI: 10.3928/1542-8877-19950501-19

关键词: Iris (anatomy)Provocative testEye diseaseMedicineUltrasound biomicroscopyMeasurement methodThickeningOpticsDark roomUltrasound

摘要: Ultrasound biomicroscopy can image the relationship of angle structures under any lighting conditions. Eight patients with narrow angles were examined in light and dark. All eyes showed iris thickening shortening, increased anterior convexity iris, varying degrees narrowing In one eye, was completely closed. Iridotomy this patient flattened profile opened angle. The degree opening observed after iridotomy did not change depending on allows imaging dynamic changes ocular as they occur provides information regarding occludability

参考文章(5)
A. C. Higgitt, The Dark-Room Test British Journal of Ophthalmology. ,vol. 38, pp. 242- 247 ,(1954) , 10.1136/BJO.38.4.242
J Gloster, D Poinoosawmy, Changes in intraocular pressure during and after the dark-room test. British Journal of Ophthalmology. ,vol. 57, pp. 170- 178 ,(1973) , 10.1136/BJO.57.3.170
Charles J. Pavlin, Kasia Harasiewicz, Michael D. Sherar, F. Stuart Foster, CLINICAL USE OF ULTRASOUND BIOMICROSCOPY Ophthalmology. ,vol. 98, pp. 287- 295 ,(1991) , 10.1016/S0161-6420(91)32298-X
Charles J. Pavlin, Michael D. Sherar, F. Stuart Foster, Subsurface Ultrasound Microscopic Imaging of the Intact Eye Ophthalmology. ,vol. 97, pp. 244- 250 ,(1990) , 10.1016/S0161-6420(90)32598-8
Ragnar Törnquist, Dark-room test on eyes with a shallow anterior chamber. Acta Ophthalmologica. ,vol. 36, pp. 664- 671 ,(2009) , 10.1111/J.1755-3768.1958.TB02276.X