Effect of replacement of vanadium by iron on the electrochemical behaviour of titanium alloys in simulated physiological media

作者: D. Mareci , V. Lucero , J. Mirza

DOI: 10.3989/REVMETALM.0750

关键词: MetallurgyVanadiumCorrosionTitanium alloyDielectric spectroscopyElectrochemistryPolarization (electrochemistry)High impedanceOpen-circuit voltageAnalytical chemistryMaterials science

摘要: The electrochemical behaviour of Ti6Al4V, Ti6Al3.5Fe and Ti5Al2.5Fe alloys has been evaluated in Ringer’s solution at 25 °C. effect the substitution vanadium Ti6Al4V alloy specifically addressed. evaluation corrosion resistance was carried out through analysis open circuit potential variation with time, potentiodynamic polarization curves, impedance spectroscopy (EIS) tests. Very low current densities were obtained (order nA/cm 2 ) from curves EIS, indicating a typical passive for all investigated alloys. EIS results exhibited relative capacitive (large resistance) phase angle close to –80° high values 105 Ω•cm medium frequencies, which are indicative formation highly stable film on these solution. In conclusion, is not affected by iron.

参考文章(21)
Eliades T, Passive film growth on titanium alloys: physicochemical and biologic considerations. International Journal of Oral & Maxillofacial Implants. ,vol. 12, pp. 621- ,(1997)
Sû. Canay, N. Hersek, A. Çulha, S. Bilgiç, EVALUATION OF TITANIUM IN ORAL CONDITIONS AND ITS ELECTROCHEMICAL CORROSION BEHAVIOUR Journal of Oral Rehabilitation. ,vol. 25, pp. 759- 764 ,(1998) , 10.1046/J.1365-2842.1998.00302.X
S. P. Kedici, A. Abbas Aksüt, M. Ali Kílíçarslan, G. Bayramog Lu, K. Gökdemir, Corrosion behaviour of dental metals and alloys in different media Journal of Oral Rehabilitation. ,vol. 25, pp. 800- 808 ,(1998) , 10.1046/J.1365-2842.1998.00305.X
Tae-In Kim, Jun-Hyun Han, In-Seok Lee, Kyu-Hwan Lee, Myung-Chul Shin, Boo-Byung Choi, New titanium alloys for biomaterials: a study of mechanical and corrosion properties and cytotoxicity. Bio-medical Materials and Engineering. ,vol. 7, pp. 253- 263 ,(1997) , 10.3233/BME-1997-7404
M. Stern, A. L. Geaby, Electrochemical Polarization I . A Theoretical Analysis of the Shape of Polarization Curves Journal of The Electrochemical Society. ,vol. 104, pp. 56- 63 ,(1957) , 10.1149/1.2428496
C.R.M. Afonso, G.T. Aleixo, A.J. Ramirez, R. Caram, Influence of cooling rate on microstructure of Ti–Nb alloy for orthopedic implants Materials Science and Engineering: C. ,vol. 27, pp. 908- 913 ,(2007) , 10.1016/J.MSEC.2006.11.001
Sérgio Luiz de Assis, Stephan Wolynec, Isolda Costa, Corrosion characterization of titanium alloys by electrochemical techniques Electrochimica Acta. ,vol. 51, pp. 1815- 1819 ,(2006) , 10.1016/J.ELECTACTA.2005.02.121
M.A. Khan, R.L. Williams, D.F. Williams, The corrosion behaviour of Ti-6Al-4V, Ti-6Al-7Nb and Ti-13Nb-13Zr in protein solutions. Biomaterials. ,vol. 20, pp. 631- 637 ,(1999) , 10.1016/S0142-9612(98)00217-8
T.M. Silva, J.E. Rito, A.M.P. Simões, M.G.S. Ferreira, M. da Cunha Belo, K.G. Watkins, Electrochemical characterisation of oxide films formed on Ti6A14V alloy implanted with Ir for bioengineering applications Electrochimica Acta. ,vol. 43, pp. 203- 211 ,(1998) , 10.1016/S0013-4686(97)00226-0
I. C. Lavos-Valereto, I. Costa, S. Wolynec, The electrochemical behavior of Ti-6Al-7Nb alloy with and without plasma-sprayed hydroxyapatite coating in Hank's solution. Journal of Biomedical Materials Research. ,vol. 63, pp. 664- 670 ,(2002) , 10.1002/JBM.10351