作者: Harutyun Melikyan , Emma Danielyan , Seungwan Kim , Jongchel Kim , Arsen Babajanyan
DOI: 10.1016/J.MEDENGPHY.2011.07.020
关键词: Non invasive 、 Microwave reflection 、 Materials science 、 Biochemistry 、 Microwave sensor 、 Aqueous solution 、 Microwave 、 Resonator 、 D-Glucose 、 In vitro 、 Chromatography
摘要: We have developed an electromagnetic resonant spiral sensor and measured the glycemia in pig blood concentration of D-glucose aqueous solution by using a real-time interaction phenomenon between microwave liquid. could determine glucose with minimal resolution 5 mg/dl 100-600 range at operating frequencies about 7.65 GHz (for solution) 7.77 sample). The change brings changes reflection coefficient due to resonator solution. vitro results show signal-to-noise ratio 34 dB, minimum detectible signal level 0.022 dB/(mg/dl). Our proposed system provides unique approach for non-invasive non-contact monitoring, it may serve as bloodless glucometer.