作者: Lei Ge , Shoumei Wang , Xianrang Song , Shenguang Ge , Jinghua Yu
DOI: 10.1039/C2LC40325K
关键词: Microfluidics 、 Immunoassay 、 Multiplexing 、 Paper based 、 Chemiluminescence 、 Chemistry 、 Ag nanoparticles 、 Chemiluminescence immunoassay 、 Chromatography 、 Multiplex
摘要: A novel 3D microfluidic paper-based immunodevice, integrated with blood plasma separation from whole samples, automation of rinse steps, and multiplexed CL detections, was developed for the first time based on principle origami (denoted as origami-based device). This device, comprised one test pad surrounded by four folding tabs, could be patterned fabricated wax-printing paper in bulk. In this work, a sandwich-type chemiluminescence (CL) immunoassay introduced into which separate operational procedures several steps including (i) pads above/below (ii) addition reagent/buffer under specific sequence. The behavior, separation, washing protocol, incubation were investigated work. combined typical luminuol-H2O2 system catalyzed Ag nanoparticles, showed excellent analytical performance simultaneous detection tumor markers. samples assayed results obtained agreement reference values parallel single-analyte test. provides new strategy low-cost, sensitive, multiplex point-of-care diagnostics.