Method of analyzing medical specimens with improved length of analytical run determination

作者: James O. Westgard , Paul Mountain

DOI:

关键词: StatisticsControl materialVolume (compression)EngineeringOperations management

摘要: In high volume laboratory analysis of patient medical specimens, control materials are inserted at periodical intervals (e.g. once per eight hours) and analyzed. If the results deviate from mean according to statistical rules, run analyses prior insertion has in past been rejected, often causing a substantial loss because were not sufficiently frequently. According invention test is determined, when deviates predetermined by more than selected amount, this determination used, reject run, but rather trigger materials. shows that control, made thereto reported resumed. material out then rejected.

参考文章(6)
C G Fraser, P H Petersen, R Haeckel, C Ricos, Proposed quality specifications for the imprecision and inaccuracy of analytical systems for clinical chemistry. European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies. ,vol. 30, pp. 311- 317 ,(1992)
George S. Cembrowski, Elliot P. Chandler, James O. Westgard, Assessment of "Average of Normals" quality control procedures and guidelines for implementation. American Journal of Clinical Pathology. ,vol. 81, pp. 492- 499 ,(1984) , 10.1093/AJCP/81.4.492
J O Westgard, P H Petersen, D A Wiebe, Laboratory process specifications for assuring quality in the U.S. National Cholesterol Education Program. Clinical Chemistry. ,vol. 37, pp. 656- 661 ,(1991) , 10.1093/CLINCHEM/37.5.656
J O Westgard, P L Barry, M R Hunt, T Groth, A multi-rule Shewhart chart for quality control in clinical chemistry Clinical Chemistry. ,vol. 27, pp. 493- 501 ,(1981) , 10.1093/CLINCHEM/27.3.493