Acquistion and Accuracy in Rapid NMR Imaging Methods

作者: Donald B. Twieg

DOI: 10.1002/MRM.1910020504

关键词: Sampling (signal processing)WaveformNuclear magnetic resonanceAlgorithmMathematicsNoise powerComputationTransfer functionFourier transformObserver (quantum physics)Signal

摘要: The echo-planar imaging method (EPI) and its variants can acquire data sufficient to form an image in a much shorter time than the Fourier or projection-reconstruction methods. Fundamental requirements on gradient waveforms signal sampling are discussed for EPI SEPI (a variant using sinusoidal waveforms), based k-trajectory formulation of acquisition NMR imaging. Using new simulation method, defects demonstrated which result from failure meet these requirements. Faithful rendition theoretically ideal places severe demands system response rapid It is shown, however, that nonideal need not degrade images if appropriate reconstruction computations used. Transfer functions noise power spectra derived SEPI, ratio two used derive observer signal-to-noise ratios simple visual interpretive tasks. shown performance differ significantly between some tasks, same system.

参考文章(16)
George M. Siouris, D. L. Cohn, J. L. Melsa, Decision and estimation theory ,(1978)
L. F. Feiner, P. R. Locher, On NMR spin imaging by magnetic field modulation Applied physics. ,vol. 22, pp. 257- 271 ,(1980) , 10.1007/BF00899875
Kevin F. King, Paul R. Moran, A unified description of NMR imaging, data-collection strategies, and reconstruction Medical Physics. ,vol. 11, pp. 1- 14 ,(1984) , 10.1118/1.595470
Andrew P. Sage, James L. Melsa, W. J. Steinway, Estimation theory with applications to communications and control ,(1979)
Stig Ljunggren, A simple graphical representation of fourier-based imaging methods Journal of Magnetic Resonance (1969). ,vol. 54, pp. 338- 343 ,(1983) , 10.1016/0022-2364(83)90060-4
A. Burgess, R. Wagner, R. Jennings, H. Barlow, Efficiency of human visual signal discrimination. Science. ,vol. 214, pp. 93- 94 ,(1981) , 10.1126/SCIENCE.7280685
Edwin D. Becker, James A. Feretti, Thomas C. Farrar, Driven equilibrium Fourier transform spectroscopy. A new method for nuclear magnetic resonance signal enhancement. Journal of the American Chemical Society. ,vol. 91, pp. 7784- 7785 ,(1969) , 10.1021/JA50001A068
R. R. Ernst, W. A. Anderson, Application of Fourier Transform Spectroscopy to Magnetic Resonance Review of Scientific Instruments. ,vol. 37, pp. 93- 102 ,(1966) , 10.1063/1.1719961