Determination of serum ionised calcium by ion-exchange electrode in normal subjects

作者: S.C. Conceicao , M.K. Ward , F. Alvarez-Ude , P. Aljama , P. Smith

DOI: 10.1016/0009-8981(78)90127-4

关键词: MoleMean valueIon exchangeNormal peopleCalciumWhole bloodChromatographyElectrodeChemistryIonised calcium

摘要: Abstract Serum ionised calcium concentration [Ca 2+ ] was measured with a selective electrode in 65 normal people. A mean value of 1.06 mmol/1 (± 0.04 S.D.) an actual range 0.97 to 1.13 obtained. samples refrigerated at 4° C for 24 h were satisfactory analysis. Storage whole blood 6 room temperature before separation caused very small error (+0.04 mmol/1). Mean results and ranges similar males females there no significant correlation age. No change serum found following meal. total showed slight ( r = 0.35). The method is reproducible sensitive.

参考文章(19)
Franklin C. McLean, A. Baird Hastings, A BIOLOGICAL METHOD FOR THE ESTIMATION OF CALCIUM ION CONCENTRATION Journal of Biological Chemistry. ,vol. 107, pp. 337- 350 ,(1934) , 10.1016/S0021-9258(18)75414-8
H Husdan, M Leung, D Oreopoulos, A Rapoport, Measurement of serum and plasma ionic calcium with the "Space-Stat 20 Ionized Calcium Analyzer". Clinical Chemistry. ,vol. 23, pp. 1775- 1777 ,(1977) , 10.1093/CLINCHEM/23.9.1775
A M Parfitt, Letter: Correction of plasma calcium measurements. BMJ. ,vol. 1, pp. 520- 520 ,(1974) , 10.1136/BMJ.1.5906.520
James C. Low, Ionic Calcium Determination in Primary Hyperparathyroidism JAMA: The Journal of the American Medical Association. ,vol. 223, pp. 152- 155 ,(1973) , 10.1001/JAMA.1973.03220020022005
Irwin Oreskes, Carl Hirsch, Kathryn S. Douglas, Sherman Kupfer, Measurement of ionized calcium in human plasma with a calcium selective electrode. Clinica Chimica Acta. ,vol. 21, pp. 303- 313 ,(1968) , 10.1016/0009-8981(68)90060-0