作者: Z. L. Chen , M. A. Adams
DOI: 10.1007/BF02467748
关键词: Magnesium 、 Chromatography 、 Chemistry 、 Conductometry 、 Elution 、 Ion chromatography 、 Conductivity 、 Magnesium ion 、 Column chromatography 、 Detection limit
摘要: 2,6-Pyridinedicarboxylic acid (2,6-PDC) was investigated as an eluent for ion chromatographic (IC) separation of inorganic anions, Mg2+ and Ca2+ followed by UV conductivity detection. Other eluents containing trimesic, pyromellitic commonly used gluconate/borate were also studied. Eluents 2, 6-PDC gave reasonable resolution detection sensitivity all analytes interest. The effects pH concentration 2,6-PDC on the retention systematically. HCO3−, H2PO4−, Cl−, NO2−, NO3−, SO4−, within 11 minutes performed using 1 mM at 6.0. Detection limits in range 5–15 μM 1.5–4 proposed method offers a quick, sensitive simple means determination common particulaly water samples.