Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in patients with glaucoma.

作者: Evelien Vandewalle , Sofie Vandenbroeck , Ingeborg Stalmans , Thierry Zeyen

DOI: 10.1177/112067210901900516

关键词:

摘要: PURPOSE To compare the intraocular pressure (IOP) readings taken by ICare, Pascal dynamic contour tonometer (DCT), and ocular response analyzer (ORA) with those Goldmann applanation (GAT). We sought to evaluate influence of central corneal thickness (CCT) on IOP measurements patients' preferences for four tonometers. METHODS In this prospective study, 93 eyes from patients were examined. Patients randomly divided into 4 groups vary order in which tonometers applied. CCT was measured an ultrasound pachymeter. RESULTS The average 558+/-47.4 microm. mean +/- standard deviation GAT, DCT, ORA (Goldmann-correlated IOP) (ORA(GC)) 15.1+/-4.8, 15.7+/-5.7, 18.2+/-5.1, 18.3+/-6.6 mmHg, respectively. There no significant difference between obtained GAT ICare (p=0.14). also levels DCT (p=0.26). correlation instruments. Bland-Altman graphs showed disagreement other preference among instruments (p=0.48). CONCLUSIONS consistent whereas correspond well ORA(GC) measurements. both overestimated readings.

参考文章(32)
Michael J Doughty, Mohammed L Zaman, Human Corneal Thickness and Its Impact on Intraocular Pressure Measures Survey of Ophthalmology. ,vol. 44, pp. 367- 408 ,(2000) , 10.1016/S0039-6257(00)00110-7
Dimitrios S Siganos, Georgios I Papastergiou, Carlos Moedas, Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. Journal of Cataract and Refractive Surgery. ,vol. 30, pp. 746- 751 ,(2004) , 10.1016/J.JCRS.2003.12.033
Nima Pakrou, Tim Gray, Richard Mills, John Landers, Jamie Craig, Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry Journal of Glaucoma. ,vol. 17, pp. 43- 47 ,(2008) , 10.1097/IJG.0B013E318133FB32
Makoto Nakamura, Urtogtah Darhad, Yasuko Tatsumi, Miyuki Fujioka, Azusa Kusuhara, Hidetaka Maeda, Akira Negi, Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. American Journal of Ophthalmology. ,vol. 142, pp. 332- 334 ,(2006) , 10.1016/J.AJO.2006.02.035
Inci Koçak, Selim Orgül, Ayhan Saruhan, Ivan Haefliger, Phillip Hendrickson, Josef Flammer, Measurement of Intraocular Pressure with a Modern Noncontact Tonometer Ophthalmologica. ,vol. 212, pp. 81- 87 ,(1998) , 10.1159/000027284
Paolo Brusini, Maria Letizia Salvetat, Marco Zeppieri, Claudia Tosoni, Lucia Parisi, Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. Journal of Glaucoma. ,vol. 15, pp. 213- 217 ,(2006) , 10.1097/01.IJG.0000212208.87523.66
Hartmut E Kanngiesser, Christoph Kniestedt, Yves C A Robert, Dynamic contour tonometry: presentation of a new tonometer. Journal of Glaucoma. ,vol. 14, pp. 344- 350 ,(2005) , 10.1097/01.IJG.0000176936.16015.4E
Antti Kontiola, A new electromechanical method for measuring intraocular pressure. Documenta Ophthalmologica. ,vol. 93, pp. 265- 276 ,(1997) , 10.1007/BF02569066
Aachal Kotecha, Ahmed Elsheikh, Cynthia R. Roberts, Haogang Zhu, David F. Garway-Heath, Corneal Thickness- and Age-Related Biomechanical Properties of the Cornea Measured with the Ocular Response Analyzer Investigative Ophthalmology & Visual Science. ,vol. 47, pp. 5337- 5347 ,(2006) , 10.1167/IOVS.06-0557