作者: Xia Zhao , Zilong Zhao , Qufei Shi , Meng Dou , Rui Zheng
DOI: 10.1109/PHM-JINAN48558.2020.00106
关键词:
摘要: Dry block furnaces are widely used in on-site measurement and calibration of food, petrochemical, pharmaceutical, semiconductor, other manufacturing enterprises. The accuracy dry-block furnace temperature has always been a problem for calibrator. research on the mainly focuses stability measurement, method, distribution uniformity so on. Therefore, axial is an important factor affecting its accuracy. In this study, (Const 660 produced China) selected as object to study relationship between dry well by experimental method. results found that from bottom wellhead side, probe insertion depth should be close possible well, closer ambient temperature, higher measurement. provides more comprehensive reliable data application field calibration, improve better furnace.