作者: Fengyi Li , Shaoyi Shi , Weigang Ma , Xing Zhang
DOI: 10.1063/1.5064426
关键词:
摘要: A method involving a vibrating hot wire is proposed for measuring the viscosity and thermal conductivity of liquids. platinum bent into semicircular shape immersed in sample liquid presence static magnetic field. Alternating current then applied to wire, causing it vibrate generate heat. At low frequency, frequency response vibration used calculate viscosity. high amplitude less than molecular free path, obtained from temperature dependence resistance. The validated using water, toluene, anhydrous ethanol, ethanediol as test samples. measurement uncertainty estimated be 1.5% (k = 1) 0.7% 2)