Evaluating the impact of x‐ray spectral shape on image quality in flat‐panel CT breast imaging

作者: Stephen J. Glick , Samta Thacker , Xing Gong , Bob Liu

DOI: 10.1118/1.2388574

关键词:

摘要: In recent years, there has been an increasing interest in exploring the feasibility of dedicated computed tomography (CT) breast imaging using a flat-panel digital detector truncated cone-beam geometry. Preliminary results are promising and it appears as if three-dimensional tomographic great potential for reducing masking effect superimposed parenchymal structure typically observed with conventional mammography. this study, mathematical framework used determining optimal design acquisition parameters such CT system is described. The ideal observer signal-to-noise ratio (SNR) figure merit, under assumptions that linear shift invariant. Computation SNR parallel-cascade model to predict signal noise propagation through detector, well realistic lesion detection task imaging. For all evaluations, total mean glandular dose study was constrained be approximately equivalent two-view mammography study. presented explore x-ray spectral shape across extensive range kVp settings, filter material types, thicknesses. give indication how can affect image quality

参考文章(46)
Wei Zhao, Dylan C. Hunt, Kenkichi Tanioka, John A. Rowlands, Indirect flat-panel detector with avalanche gain Medical Imaging 2004: Physics of Medical Imaging. ,vol. 5368, pp. 150- 161 ,(2004) , 10.1117/12.536863
R. J. Jennings, R. J. Eastgate, M. P. Siedband, D. L. Ergun, Optimal x-ray spectra for screen-film mammography Medical Physics. ,vol. 8, pp. 629- 639 ,(1981) , 10.1118/1.595021
Angel R. Pineda, Harrison H. Barrett, Figures of merit for detectors in digital radiography. I. Flat background and deterministic blurring. Medical Physics. ,vol. 31, pp. 348- 358 ,(2004) , 10.1118/1.1631426
John M. Boone, J. Anthony Seibert, An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV. Medical Physics. ,vol. 24, pp. 1661- 1670 ,(1997) , 10.1118/1.597953
Ruola Ning, Yong Yu, David L. Conover, Xianghua Lu, Huiguang He, Zikuan Chen, Linda Schiffhauer, Jeanne Cullinan, Preliminary System Characterization of Flat Panel Detector-based Cone Beam CT for Breast Imaging Medical Imaging 2004: Physics of Medical Imaging. ,vol. 5368, pp. 292- 303 ,(2004) , 10.1117/12.536231
J. M. Boone, N. Shah, T. R. Nelson, A comprehensive analysis of DgN(CT) coefficients for pendant-geometry cone-beam breast computed tomography. Medical Physics. ,vol. 31, pp. 226- 235 ,(2004) , 10.1118/1.1636571
Martin P. Tornai, Randolph L. McKinley, Caryl N. Bryzmialkiewicz, Priti Madhav, Spencer J. Cutler, Dominic J. Crotty, James E. Bowsher, Ehsan Samei, Carey E. Floyd, Jr., Design and development of a fully 3D dedicated x-ray computed mammotomography system Medical Imaging 2005: Physics of Medical Imaging. ,vol. 5745, pp. 189- 197 ,(2005) , 10.1117/12.595636
A. Karellas, N. Tan, M. A. Davis, C. J. D'Orsi, Absorption And Scattering Properties Of Elemental Filters For X-Ray Beams Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems (PACS IV) for Medical Applications. ,vol. 626, pp. 170- 175 ,(1986) , 10.1117/12.975389
Biao Chen, Ruola Ning, Cone-beam volume CT breast imaging: feasibility study. Medical Physics. ,vol. 29, pp. 755- 770 ,(2002) , 10.1118/1.1461843