Precision X-Ray Diffractometry Using Powder Specimens

作者: L. F. Vassamillet , H. W. King

DOI: 10.1007/978-1-4684-8783-1_12

关键词:

摘要: Errors concerned with the geometry of counter tube diffractometer method, nature x-ray source, and technique collecting data are reviewed briefly. arising from imperfections in instrument misalignment source respect to were studied experimentally. The results presented discussed terms resultant error determination lattice parameter a cubic crystal. determined both analytically empirically relation extrapolation procedures commonly used for diffractometers. It is shown that, depending on construction instrument, effect gears may almost double final extrapolated value parameter. (P.C.H.)

参考文章(17)
J. Ladell, W. Parrish, J. Taylor, Center-of-gravity method of precision lattice parameter determination Acta Crystallographica. ,vol. 12, pp. 253- 254 ,(1959) , 10.1107/S0365110X59000743
A. J. C. Wilson, On the correction of lattice spacings for refraction Mathematical Proceedings of the Cambridge Philosophical Society. ,vol. 36, pp. 485- 489 ,(1940) , 10.1017/S0305004100017539
A. J. C. Wilson, Some problems in the definition of wavelengths in x-ray crystallography* Zeitschrift Fur Kristallographie. ,vol. 111, pp. 471- 476 ,(1959) , 10.1524/ZKRI.1959.111.1-6.471
W. L. Bond, Precision lattice constant determination Acta Crystallographica. ,vol. 13, pp. 814- 818 ,(1960) , 10.1107/S0365110X60001941
H. W. King, L. F. Vassamillet, PRECISION LATTICE PARAMETER DETERMINATION BY DOUBLE-SCANNING DIFFRACTOMETRY Advances in x-ray analysis. ,vol. 5, pp. 78- 85 ,(1961) , 10.1154/S0376030800001440
Thomas C. Furnas, Eugene W. White, New Instruments for X-Ray Analysis Advances in x-ray analysis. ,vol. 4, pp. 521- 537 ,(1960) , 10.1154/S0376030800001312
James A. Ibers, Lonsdale, Kathleen, Dame, John S. Kasper, Gerard D. Rieck, Walter C. Hamilton, Norman Fordyce McKerron Henry, Caroline H. MacGillavry, International tables for X-ray crystallography UT Back-in-Print Service. ,(1962)