Routine individualised patient dosimetry using electronic portal imaging devices

作者: Sebastiaan M.J.J.G. Nijsten , Ben J. Mijnheer , André L.A.J. Dekker , Philippe Lambin , André W.H. Minken

DOI: 10.1016/J.RADONC.2007.03.003

关键词:

摘要: Background and purpose: To analyse the results of routine EPID measurements for individualised patient dosimetry. Materials methods: Calibrated camera-based EPIDs were used to measure central field dose, which was compared with a dose prediction at level. For transit dosimetry, data calculated using transmission scatter, measured values. Furthermore, backprojected an in vivo value 5 cm depth water (D5) directly D5 from treatment planning system. Dose differences per session by weighting values number monitor units beam. Reported errors categorised analysed approximately 37,500 images 2511 patients during period 24 months. Results: Pre-treatment showed mean difference 0.0 ± 1.7% (1 SD). Transfer detected corrected prior first session. An accelerator output variation about 4% found between two weekly QC measurements. Patient dosimetry � 0.7 5.2% SD) 0.3 5.6% session, respectively. could be related set-up errors, organ motion, erroneous density corrections changes anatomy. Conclusions: can routinely accurately verify parameter transfer machine output. By applying more insight obtained respect different error sources influencing delivery patient. c 2007 Elsevier Ireland Ltd. All rights reserved. Radiotherapy Oncology 83 (2007) 65–75.

参考文章(26)
Moyed Miften, Mark Wiesmeyer, Suzanne Monthofer, Ken Krippner, Implementation of FFT convolution and multigrid superposition models in the FOCUS RTP system. Physics in Medicine and Biology. ,vol. 45, pp. 817- 833 ,(2000) , 10.1088/0031-9155/45/4/301
Wouter J. C. van Elmpt, Sebastiaan M. J. J. G. Nijsten, Robert F. H. Schiffeleers, André L. A. J. Dekker, Ben J. Mijnheer, Philippe Lambin, André W. H. Minken, A Monte Carlo based three-dimensional dose reconstruction method derived from portal dose images. Medical Physics. ,vol. 33, pp. 2426- 2434 ,(2006) , 10.1118/1.2207315
V. N. Hansen, P. M. Evans, W. Swindell, The application of transit dosimetry to precision radiotherapy Medical Physics. ,vol. 23, pp. 713- 721 ,(1996) , 10.1118/1.597719
William Swindell, Philip M. Evans, Scattered radiation in portal images: A Monte Carlo simulation and a simple physical model Medical Physics. ,vol. 23, pp. 63- 73 ,(1996) , 10.1118/1.597792
Esther Bloemen-van Gurp, Wim du Bois, Peter Visser, Iaı̈n Bruinvis, Derk Jalink, Joep Hermans, Philippe Lambin, Clinical dosimetry with MOSFET dosimeters to determine the dose along the field junction in a split beam technique Radiotherapy and Oncology. ,vol. 67, pp. 351- 357 ,(2003) , 10.1016/S0167-8140(03)00035-5
Sara C Erridge, Yvette Seppenwoolde, Sara H Muller, Marcel van Herk, Katrien De Jaeger, José S.A Belderbos, Liesbeth J Boersma, Joos V Lebesque, Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer. Radiotherapy and Oncology. ,vol. 66, pp. 75- 85 ,(2003) , 10.1016/S0167-8140(02)00287-6
Mathilda van Zijtveld, Maarten L.P. Dirkx, Hans C.J. de Boer, Ben J.M. Heijmen, 3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID. Radiotherapy and Oncology. ,vol. 82, pp. 201- 207 ,(2007) , 10.1016/J.RADONC.2006.12.010
Kasper L. Pasma, Marco Kroonwijk, Sandra Quint, Andries G. Visser, Ben J.M. Heijmen, Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients International Journal of Radiation Oncology Biology Physics. ,vol. 45, pp. 1297- 1303 ,(1999) , 10.1016/S0360-3016(99)00328-4
Marcel Van Herk, R Boellaard, M Essers, M van Herk, BJ Mijnheer, New Method to Obtain the Midplane Dose Using Portal In Vivo Dosimetry International Journal of Radiation Oncology Biology Physics. ,vol. 41, pp. 465- 474 ,(1998) , 10.1016/S0360-3016(98)00048-0
Claudio Fiorino, Daniela Corletto, Paola Mangili, Sara Broggi, Antonio Bonini, Giovanni Mauro Cattaneo, Rossella Parisi, Alberto Rosso, Patrizia Signorotto, Eugenio Villa, Riccardo Calandrino, Quality assurance by systematic in vivo dosimetry: results on a large cohort of patients. Radiotherapy and Oncology. ,vol. 56, pp. 85- 95 ,(2000) , 10.1016/S0167-8140(00)00195-X