In vivo comparison between two tip pressure transducer systems

作者: André E. Aubert , Matty Vrolix , Hilaire De Geest , Frans Van de Werf

DOI: 10.1007/BF01142487

关键词:

摘要: Experimental findings are presented of anin vivo comparison between a Sentron catheter and another tip transducer manometer: Millar microtip catheter. Both catheters have been used simultaneously in the left ventricle dogs. Pressure variations were elicited by drug infusion. values derivatives obtained from both systems compared. A cross correlation episodes two pressures was computed. Results this study showed good ventricular systolic pressure measured with manometers (R=0.992, p<0.0001), end-diastolic (R=0.809, p<0.0001) first derivatives: positive derivative (R=0.993, negative (R=0.634, p<0.0001). The mean signals 0.61±0.04. In frequency domain no statistical difference found location maxima peaks. It is concluded that manometer can be valid alternative, at reasonable price, to cheaper, though less accurate fluid filled more expensive ‘golden standard’

参考文章(8)
G. Hellige, Ventrikeldruckmessung mit konventionellen Kathetermanometersystemen Basic Research in Cardiology. ,vol. 71, pp. 319- 336 ,(1976) , 10.1007/BF01906457
H. D. Millar, L. E. Baker, A stable ultraminiature catheter-tip pressure transducer Medical & Biological Engineering. ,vol. 11, pp. 86- 89 ,(1973) , 10.1007/BF02477303
K L Gould, S Trenholme, J W Kennedy, In vivo comparison of catheter manometer systems with the catheter-tip micromanometer. Journal of Applied Physiology. ,vol. 34, pp. 263- 267 ,(1973) , 10.1152/JAPPL.1973.34.2.263
Bouchard F, Cornu C, Baculard P, Laurens P, Soulie P, Brial E, Bruits et pressions cardio-vasculaires enregistrés in situ à l'aide d'un micromanomètre. Archives Des Maladies Du Coeur Et Des Vaisseaux. ,vol. 52, ,(1959)