Evaluating commercially available spirometers.

作者: John L. Hankinson , Reed M. Gardner , Blair J. West

DOI: 10.1164/ARRD.1980.121.1.73

关键词:

摘要: This study was designed to: (1) determine the performance characteristics of available spirometers; (2) assess practicality and applicability American Thoracic Society's (ATS) Snowbird Workshop recommendations on Standardization Spirometry; (3) whether spirometer testing could be done with room air. Nineteen spirometers were tested 16 different forced vital capacity waveforms. Fourteen met ATS requirements. Three these 14 had difficulties in determining end expiration. devices requirements for expiratory volume one second. Ten 13 maximal voluntary ventilation satisfactory. The standards recommended by believed to applicable practical. methods need expanded include more patient Testing air is easier simpler and, most devices, just as effective heated humidif...

参考文章(6)
Henry W. Glindmeyer, Sharon T. Anderson, John E. Diem, Hans Weill, A Comparison of the Jones and Stead-Wells Spirometers Chest. ,vol. 73, pp. 596- 602 ,(1978) , 10.1378/CHEST.73.5.596
Arend Bouhuys, James A. Virgulto, Calibration of flow-volume curves Lung. ,vol. 155, pp. 123- 130 ,(1978) , 10.1007/BF02730686
David E. Shanks, James F. Morris, Clinical comparison of two electronic spirometers with a water-sealed spirometer. Chest. ,vol. 69, pp. 461- 466 ,(1976) , 10.1378/CHEST.69.4.461
Leonard D. Hudson, Clinical Evaluation of New Office Spirometer JAMA: The Journal of the American Medical Association. ,vol. 240, pp. 2754- 2755 ,(1978) , 10.1001/JAMA.1978.03290250058031
Welch Mh, Hankinson J, Francis Pb, Is spirometry practical in your office Patient Care. ,vol. 12, pp. 136- ,(1978)
Solbert Permutt, Edward Chester, William Anderson, David Cugell, Thomas L. Petty, John T. Sharp, Joseph Ross, Alfred Soffer, Dale E. Braddy, Donald F. Tierney, Office Spirometry in Clinical Practice: Statement of the American College of Chest Physicians' Committee on Clinic and Office Pulmonary Function Testing Chest. ,vol. 74, pp. 298- ,(1978) , 10.1378/CHEST.74.3.298