作者: Frederico Garcia Pinto , Daniel Andrada , Ulisses Vilela Rey , Liliam M. Silva Ansaloni , José Bento Borba da Silva
DOI: 10.1080/00032710600822890
关键词:
摘要: In the present study, determination of copper in human serum by electrothermal atomic absorption spectrometry was investigated with susceptible substances to act as chemical modifiers. The optimization method involved studies pyrolysis and atomization temperature curves diluted serum. Ruthenium best modifier. With this modifier, recoveries spiked samples on different days ranged from 92% 105% a relative standard deviation (RSD) lower than 10% tube lifetime higher 2000 cycles.