作者: Julia Transfiguracion , Alice Bernier , Normand Arcand , Parminder Chahal , Amine Kamen
DOI: 10.1016/S0378-4347(01)00330-9
关键词:
摘要: Abstract An anion-exchange–high-performance liquid chromatography (AE–HPLC) method for the quantification of adenovirus type 5 (Ad5) total particles was validated according to performance criteria precision, specificity, linearity calibration and range, limit detection, quantification, accuracy recovery. The viral were detected by absorbance at 260 nm using photodiode array detector (PDA). Cesium chloride (CsCl) purified Ad5 lysate samples used validation method. Relative standard deviations (RSDs) inter-day, intra-day precision reproducibility both less than 10 2% peak area retention time, respectively. specific which eluted 8.0 min. presence DNA does not affect recovery accurate quantification. Based on error in prediction be 10%, working range established between 2×1010 7×1011 VP/ml with correlation coefficient 0.99975, deviation 6.14×109 a slope 3.04×105 VP/ml. varied 88 106% all investigated is statistically similar 100% 95% confidence interval.