作者: Shigeru Suzuki , Shigeru Furui , Ichiro Yamaguchi , Masafumi Yamagishi , Toshi Abe
DOI: 10.1016/J.JVIR.2008.05.020
关键词:
摘要: Purpose To measure the entrance surface dose (ESD) during three-dimensional (3D) imaging of a phantom and evaluate method to estimate ESD with use dose–area product (DAP) values. Materials Methods The present study used an angiographic unit digital flat-panel system for 3D imaging. DAP values ESDs were evaluated on square phantoms thicknesses 12.0, 15.0, 18.0, 21.0, 25.0 cm 5-second acquisitions. measured lateral posterior surfaces phantom. also human-shaped various table heights, centerline, right middle axillary line, their midpoint. Results 7.3 mGy, 12.1 18.8 26.9 38.5 mGy 12 cm, 15 18 21 25 respectively. correlated ( r = 0.998, P Conclusions single was considerably lower than thresholds radiation skin injuries. are useful maximum patient