作者: Robert T. Kester , Tomasz S. Tkaczyk , Michael R. Descour , Todd Christenson , Rebecca Richards-Kortum
DOI: 10.1364/OE.15.002409
关键词:
摘要: A disposable high numerical aperture microendoscope objective has been designed, fabricated, and tested for use with a fiber confocal reflectance microscope. The uses precision LIGA fabricated components to integrate imaging hydraulic suction lines into housing that measures only 3.85 mm in outer diameter 14.65 length. hydraulics are used translate tissue through the focal plane three dimensional imaging. This device is diffraction limited λ = 850 nm, of 1.0, field view 250 μm, working distance 450 μm. intended vivo precancerous cells.