Proximal Component Modularity in THA—At What Cost?: An Implant Retrieval Study

作者: A. M. Kop , C. Keogh , E. Swarts

DOI: 10.1007/S11999-011-2155-9

关键词:

摘要: Background While modular femoral heads have been used in THA for decades, a recent innovation is second neck-stem taper junction. Clinical advantages include intraoperative adjustment of leg length, anteversion, and easier revision, all providing flexibility to the surgeon; however, there reports catastrophic fracture, cold welding, corrosion fretting

参考文章(25)
Alan M. Kop, Eric Swarts, Corrosion of a Hip Stem With a Modular Neck Taper Junction The Journal of Arthroplasty. ,vol. 24, pp. 1019- 1023 ,(2009) , 10.1016/J.ARTH.2008.09.009
JOSHUA J. JACOBS, ANASTASIA K. SKIPOR, LESLIE M. PATTERSON, NADIM J. HALLAB, WAYNE G. PAPROSKY, JONATHAN BLACK, JORGE O. GALANTE, Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. Journal of Bone and Joint Surgery, American Volume. ,vol. 80, pp. 1447- 1458 ,(1998) , 10.2106/00004623-199810000-00006
M. Schramm, D. C. Wirtz, U. Holzwarth, R. P. Pitto, The Morse taper junction in modular revision hip replacement--a biomechanical and retrieval analysis. Biomedizinische Technik. ,vol. 45, pp. 105- 109 ,(2000) , 10.1515/BMTE.2000.45.4.105
M. Viceconti, O. Ruggeri, A. Toni, A. Giunti, Design-related fretting wear in modular neck hip prosthesis Journal of Biomedical Materials Research. ,vol. 30, pp. 181- 186 ,(1996) , 10.1002/(SICI)1097-4636(199602)30:2<181::AID-JBM7>3.0.CO;2-N
Eduardo A. Salvati, Jay R. Lieberman, Olga L. Huk, Brian G. Evans, Complications of femoral and acetabular modularity. Clinical Orthopaedics and Related Research. pp. 85- 93 ,(1995) , 10.1097/00003086-199510000-00009
Francesco Pallini, Luca Cristofolini, Francesco Traina, Aldo Toni, Modular hip stems: determination of disassembly force of a neck-stem coupling. Artificial Organs. ,vol. 31, pp. 166- 170 ,(2007) , 10.1111/J.1525-1594.2007.00359.X
D.C. Fricker, R. Shivanatii, Fretting corrosion studies of universal femoral head prostheses and cone taper spigots Biomaterials. ,vol. 11, pp. 495- 500 ,(1990) , 10.1016/0142-9612(90)90064-W
Jay R. Goldberg, Jeremy L. Gilbert, Joshua J. Jacobs, Thomas W. Bauer, Wayne Paprosky, Sue Leurgans, A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clinical Orthopaedics and Related Research. ,vol. 401, pp. 149- 161 ,(2002) , 10.1097/00003086-200208000-00018
M.A. Khan, R.L. Williams, D.F. Williams, In-vitro corrosion and wear of titanium alloys in the biological environment Biomaterials. ,vol. 17, pp. 2117- 2126 ,(1996) , 10.1016/0142-9612(96)00029-4
Scott M. Sporer, Craig DellaValle, Joshua Jacobs, Markus Wimmer, A Case of Disassociation of a Modular Femoral Neck Trunion After Total Hip Arthroplasty The Journal of Arthroplasty. ,vol. 21, pp. 918- 921 ,(2006) , 10.1016/J.ARTH.2005.10.014