摘要: Publisher Summary The most successful integration of megavoltage computed tomography (MVCT) imaging with radiation therapy devices is the TomoTherapy™ Hi-ART. It currently only type patient treatment in which MVCT integrated as part device itself. Hi-ART II system can generate CT images from same X-ray beam it uses for treatment. These offer verification position prior to and potentially during therapy. amount integral dose (total tissue) that receives dependent on slice thickness total length imaged region. need be registered kVCT order establish proper positioning. Once are acquired reconstructed, therapist (RTT) has option use either automatic or manual registration, data set primary set. Axial, sagittal, coronal views available console assist operator registration process. There three different scanning modes—coarse, normal, fine—which correspond image thicknesses 6.0, 4.0, 2.0 mm, respectively.

参考文章(19)
C J Langer, Concurrent chemoradiation using paclitaxel and carboplatin in locally advanced non-small cell lung cancer. Seminars in Radiation Oncology. ,vol. 9, pp. 108- 116 ,(1999)
Robert Jeraj, Thomas R. Mackie, John Balog, Gustavo Olivera, Dave Pearson, Jeff Kapatoes, Ken Ruchala, Paul Reckwerdt, Radiation characteristics of helical tomotherapy. Medical Physics. ,vol. 31, pp. 396- 404 ,(2004) , 10.1118/1.1639148
K.M Langen, D.T.L Jones, Organ motion and its management International Journal of Radiation Oncology Biology Physics. ,vol. 50, pp. 265- 278 ,(2001) , 10.1016/S0360-3016(01)01453-5
William Swindell, Robert G. Simpson, James R. Oleson, Ching-Tai Chen, Elmer A. Grubbs, Computed tomography with a linear accelerator with radiotherapy applications Medical Physics. ,vol. 10, pp. 416- 420 ,(1983) , 10.1118/1.595391
K M Langen, S L Meeks, D O Poole, T H Wagner, T R Willoughby, P A Kupelian, K J Ruchala, J Haimerl, G H Olivera, The use of megavoltage CT (MVCT) images for dose recomputations Physics in Medicine and Biology. ,vol. 50, pp. 4259- 4276 ,(2005) , 10.1088/0031-9155/50/18/002
K J Ruchala, G H Olivera, J M Kapatoes, E A Schloesser, P J Reckwerdt, T R Mackie, Megavoltage CT image reconstruction during tomotherapy treatments Physics in Medicine and Biology. ,vol. 45, pp. 3545- 3562 ,(2000) , 10.1088/0031-9155/45/12/303
William Swindell, A 4-MV CT scanner for radiation therapy; spectral properties of the therapy beam Medical Physics. ,vol. 10, pp. 347- 351 ,(1983) , 10.1118/1.595280
Sanford L. Meeks, Joseph F. Harmon, Katja M. Langen, Twyla R. Willoughby, Thomas H. Wagner, Patrick A. Kupelian, Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit. Medical Physics. ,vol. 32, pp. 2673- 2681 ,(2005) , 10.1118/1.1990289
R. G. Simpson, C. T. Chen, E. A. Grubbs, W. Swindell, A 4-MV CT scanner for radiation therapy: The prototype system Medical Physics. ,vol. 9, pp. 574- 579 ,(1982) , 10.1118/1.595102
K J Ruchala, G H Olivera, E A Schloesser, T R Mackie, Megavoltage CT on a tomotherapy system Physics in Medicine and Biology. ,vol. 44, pp. 2597- 2621 ,(1999) , 10.1088/0031-9155/44/10/316