Determination of trace concentrations of antimony by the introduction of stibine into an inductively-coupled plasma for atomic emission spectrometry

作者: Taketoshi Nakahara , Noriaki Kikui

DOI: 10.1016/S0003-2670(00)82600-8

关键词:

摘要: Abstract A simple, rapid and sensitive method is described for the determination of trace concentrations antimony by inductively-coupled plasma atomic emission spectrometry with hydride generation. Hydrochloric acid (1 M) best medium stibine generation, but antimony(III) also effectively reduced to in 1 M malic or 0.5 tartaric acid, whereas antimony(V) gives no significant signal either these acids. This permits differential Sb(III) Sb(V). Most inter-element interference effects can be minimized thiourea, bur standard additions are recommended accurate determinations. Thiourea effective prereducing Sb(V) Sb(III). The detection limit 0.19 ng Sb ml−1 calibration graph linear up 100 μg ml−1. r.s.d, at 3.8 2.1%, respectively. applied copper metal speciation waste water.

参考文章(20)
Taketoshi. Nakahara, Syoji. Kobayashi, Soichiro. Musha, Determination of trace antimony in phosphoric acid by hydride generation and nondispersive flame atomic fluorescence spectrometry Analytical Chemistry. ,vol. 51, pp. 1589- 1591 ,(1979) , 10.1021/AC50045A058
Taketoshi Nakahara, Syoji Kobayashi, Sôichirô Musha, Non-dispersive atomic fluorescence spectrometry of nanogram amounts of antimony with a hydride generation technique Analytica Chimica Acta. ,vol. 101, pp. 375- 384 ,(1978) , 10.1016/S0003-2670(01)93372-0
R. K. Winge, V. J. Peterson, V. A. Fassel, Inductively Coupled Plasma-Atomic Emission Spectroscopy: Prominent Lines Applied Spectroscopy. ,vol. 33, pp. 206- 219 ,(1979) , 10.1366/0003702794925895