Common Mode Noise Cancellation for Electrically Non-Contact ECG Measurement System on a Chair

作者: Ko Keun Kim , Yong Kyu Lim , Kwang Suk Park

DOI: 10.1109/IEMBS.2005.1615828

关键词:

摘要: Electrically non-contact ECG measurement system on a chair can be applied to number of various fields for continuous health monitoring in daily life. However, the body is floated electrically this due capacitive electrodes and very sensitive external noises or motion artifacts which affect as common mode noise. In paper, driven-seat-ground circuit similar driven-right-leg proposed reduce The analysis equivalent performed output signal waveforms are compared between with ground. As results, improves significantly properties fully negative feedback

参考文章(2)
Bruce B. Winter, John G. Webster, Driven-right-leg circuit design IEEE Transactions on Biomedical Engineering. ,vol. 30, pp. 62- 66 ,(1983) , 10.1109/TBME.1983.325168
E.M. Spinelli, N.H. Martinez, M.A. Mayosky, A transconductance driven-right-leg circuit IEEE Transactions on Biomedical Engineering. ,vol. 46, pp. 1466- 1470 ,(1999) , 10.1109/10.804574