Translational and rotational forces on heart valve prostheses subjected ex vivo to a 4.7-T MR system.

作者: Maria-Benedicta Edwards , Roger J. Ordidge , David L. Thomas , Jeffrey W. Hand , Kenneth M. Taylor

DOI: 10.1002/JMRI.10201

关键词:

摘要: Purpose: To assess the magnetic field interactions on 60 heart valve prostheses subjected to a 4.7 T MR system. It addresses question of whether valves deemed safe at 1.5 may pose safety hazards as patients are exposed increased static fields.Materials and Methods: Ex vivo testing was performed evaluate translational rotational forces using previously described techniques.Results: Translational were detected 58 ranging from 0.5degrees 7.5degrees. Seven exhibited paramagnetic/weakly ferromagnetic behavior, 51 diamagnetic behavior. Rotational observed for 46 valves.Conclusions: Criteria used assessment expressed in terms suggest this study arc compatible with use these resonance (MR) systems fields up T.

参考文章(30)
E Kanal, F G Shellock, L Talagala, Safety considerations in MR imaging. Radiology. ,vol. 176, pp. 593- 606 ,(1990) , 10.1148/RADIOLOGY.176.3.2202008
P F New, B R Rosen, T J Brady, F S Buonanno, J P Kistler, C T Burt, W S Hinshaw, J H Newhouse, G M Pohost, J M Taveras, Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology. ,vol. 147, pp. 139- 148 ,(1983) , 10.1148/RADIOLOGY.147.1.6828719
J Michael Hasenkam, Steffen Ringgaard, Kim Houlind, René M Botnar, Hans Stødkilde-Jørgensen, Peter Boesiger, Erik Morre Pedersen, Prosthetic heart valve evaluation by magnetic resonance imaging. European Journal of Cardio-Thoracic Surgery. ,vol. 16, pp. 300- 305 ,(1999) , 10.1016/S1010-7940(99)00215-8
R P Klucznik, D A Carrier, R Pyka, R W Haid, Placement of a ferromagnetic intracerebral aneurysm clip in a magnetic field with a fatal outcome Radiology. ,vol. 187, pp. 855- 856 ,(1993) , 10.1148/RADIOLOGY.187.3.8497645
Emanuel Kanal, Frank G. Shellock, Magnetic Resonance: Bioeffects, Safety and Patient Management ,(1994)
Kotsakis A, Tan Kh, Jackson G, Is MRI a safe procedure in patients with coronary stents in situ International Journal of Clinical Practice. ,vol. 51, pp. 349- ,(1997)
Edwards Mb, Taylor Km, A profile of valve replacement surgery in the UK (1986-1997): a study from the UK Heart Valve Registry. Journal of Heart Valve Disease. ,vol. 8, pp. 697- 701 ,(1999)
J E Wolf, M Hassler, J F Le Bas, B Waksmann, M Coulomb, C Contamin, Effects of the magnetic field in magnetic resonance imaging on 15 tested cardiac valve prostheses Journal De Radiologie. ,vol. 67, pp. 661- 666 ,(1986)
Mitsuhiro Matsumoto, Hironori Hasegawa, Shinichi Inoue, Rie Taketani, Takeo Azuma, Kazuhiko Sato, Yasumitsu Kawagoe, The construction and verification of the compensation filter using digital values analysis with FCR system. Japanese Journal of Radiological Technology. ,vol. 53, pp. 233- ,(1997) , 10.6009/JJRT.KJ00001355228
Buerke M, Schreiber W, Dahm M, Thelen M, Kalden P, Pruefer D, Oelert H, In vitro investigation of prosthetic heart valves in magnetic resonance imaging: evaluation of potential hazards. Journal of Heart Valve Disease. ,vol. 10, pp. 410- ,(2001)