First human use of the Hemopump, a catheter-mounted ventricular assist device

作者: O.H. Frazier , Richard K. Wampler , J.Michael Duncan , Wayne E. Dear , Michael P. Macris

DOI: 10.1016/0003-4975(90)90155-Y

关键词:

摘要: The Hemopump, a catheter-mounted, temporary ventricular assist device, consists of an external electromechanical drive console and disposable, intraarterial axial-flow pump (21F). Power is transmitted percutaneously to the by flexible shaft within catheter. device positioned in left ventricle way femoral artery approach or through ascending aorta. Blood drawn from transvalvular inlet cannula pumped into As December 1988, Hemopump had successfully supported circulation 7 patients (5 men, 2 women) ranging age 44 72 years (mean age, 59 years) suffering cardiogenic shock (cardiac index less than 2.0 L/min/m2). Indications for use included failure be weaned cardiopulmonary bypass 4 patients, acute myocardial infarction 1, severe cardiac allograft rejection donor heart 1. Duration support ranged 26 113 hours (mean, 66 hours). Although 5 demonstrated transient hemolysis, none experienced infection, thrombosis, vascular injury. Hemodynamic variables improved all during device. were alive more 30 days after been discontinued, 3 these discharged hospital. On basis our initial clinical results, which does not require major surgical procedure insertion, provides effective, circulatory with potentially reversible failure.

参考文章(14)
J. Ward Kennedy, George C. Kaiser, Lloyd D. Fisher, Charles Maynard, James K. Fritz, William Myers, J. Gerard Mudd, Thomas J. Ryan, Joan Coggin, Multivariate discriminant analysis of the clinical and angiographic predictors of operative mortality from the Collaborative Study in Coronary Artery Surgery (CASS) The Journal of Thoracic and Cardiovascular Surgery. ,vol. 80, pp. 876- 887 ,(1980) , 10.1016/S0022-5223(19)37694-9
Michael E. DeBakey, Left ventricular bypass pump for cardiac assistance The American Journal of Cardiology. ,vol. 27, pp. 3- 11 ,(1971) , 10.1016/0002-9149(71)90076-2
J.Michael Duncan, O.H. Frazier, Branislav Radovancevic, Vladimir Velebit, Implantation techniques for the Hemopump The Annals of Thoracic Surgery. ,vol. 48, pp. 733- 735 ,(1989) , 10.1016/0003-4975(89)90810-2
MICHAEL C. McGEE, STEVEN M. PARNIS, TAKESHI NAKATANI, TIMOTHY MYERS, KURT DASSE, WILLIAM D. HARE, J. MICHAEL DUNCAN, VICTOR L. POIRIER, O. H. FRAZIER, Extended clinical support with an implantable left ventricular assist device. Asaio Journal. ,vol. 35, pp. 614- 615 ,(1989) , 10.1097/00002480-198907000-00145
John C. Norman, An Intracorporeal (Abdominal) Left Ventricular Assist Device Archives of Surgery. ,vol. 112, pp. 1442- 1451 ,(1977) , 10.1001/ARCHSURG.1977.01370120032003
Paul S. Freed, Tarik Wasfie, Barina Zado, Adrian Kantrowitz, Intraaortic balloon pumping for prolonged circulatory support The American Journal of Cardiology. ,vol. 61, pp. 554- 557 ,(1988) , 10.1016/0002-9149(88)90763-1
Eric D. Foster, Lloyd D. Fisher, George C. Kaiser, William O. Myers, Comparison of Operative Mortality and Morbidity for Initial and Repeat Coronary Artery Bypass Grafting: The Coronary Artery Surgery Study (CASS) Registry Experience The Annals of Thoracic Surgery. ,vol. 38, pp. 563- 570 ,(1984) , 10.1016/S0003-4975(10)62312-0