作者: U. Brunsmann , U. Sauer , S. Arba-Mosquera , T. Magnago , N. Triefenbach
DOI: 10.1016/J.INFRARED.2010.06.003
关键词:
摘要: Abstract Infrared thermography is used for evaluation of the mean temperature as a measure thermal load during corneal refractive surgery. An experimental method to determine emissivity and calibrate thermografic system presented. In case study on porcine eye two dimensional distributions with lateral resolution 170 μm line scans temporal 13 μs are discussed respect meaning temperature. Using newest generation surgery equipment it shown, that rise can be kept below 5 °C myopic laser in situ keratomileusis (LASIK) treatments corresponding an aberration-free correction −2.75 diopter.