作者: H.G. Scheibel , J. Porstendörfer
DOI: 10.1016/0021-8502(84)90005-3
关键词:
摘要: Abstract Experimental calibrations of tube and screen-type diffusion batteries are made by means monodisperse particles diameters ranging from 3.5 to 130 nm. Ultrafine Ag-aerosols generated a constant-temperature furnace. Electrostatic classification the polydisperse primary aerosol yields fractions with electric mobility particle in size range 3–200 The monodispersity test aerosols accuracy calculated mean confirmed analysis transmission-electron microscopy. results for good agreement theory, if entrance effects front face deposition considered. theory developed Y. S. Cheng 1980, which is based on filtration fan model filter.