A prototype MR insert PET detector with strip-line readout

作者: Heejong Kim , Yuexuan Hua , Daoming Xi , Qingguo Xie , Neville Eclov

DOI: 10.1109/NSSMIC.2017.8532635

关键词:

摘要: We are developing a MR insert PET detector using SiPM for small animal imaging. Our design is based on the strip-line signal readout that we have developed to efficiently handle large amount signals SiPM-based TOF PET. The stripline has an advantageous feature suitable building integrated within MR. In readout, digitization accomplished remotely from detection module, thereby front-end electronics of minimal; interference between two systems, e.g., magnetic field distortion due PET, can be minimized accordingly. feasibility PET/MR been demonstrated in our previous study. current prototype comprised 8 modules, which installed plastic cylindrical supporting structure. Each module consists 8x4 LYSO arrays (3x3x10 mm3each) and Hamamatsu S13361–3050NE-04 MPPC (4x4, 3.2 mm pitch); 32 routed 4 strip-lines so LYSO/SiPM row connected each strip-line. fed 72-ch multi-voltage threshold (MVT) digitizer, digitizes input by TDCs implemented FPGA. Experiments carried out evaluate performance, image reconstruction also performed simple shape object. paper, detail development explained, preliminary test results presented.

参考文章(0)