Spatial Profile Measurement Method of Absolute Detection Efficiency of an Infrared Heterodyne Receiver for the CO2 Laser Collective Thomson Scattering System in JT-60U

作者: Seishu Lee , Takashi Kondoh , Yukitoshi Miura

DOI: 10.1143/JJAP.41.6574

关键词:

摘要: A collective Thomson scattering (CTS) system based on CO2 laser is being developed in JT-60U to demonstrate the feasibility of ion temperature and alpha particle measurement international thermonuclear experimental reactor (ITER). new effective method measuring spatial profile absolute detection efficiency an infrared heterodyne receiver has been for CTS system. The viewing area divergence field view were first investigated with this method. Consequently, spot size could be narrowed 0.74 mrad by adjusting focal length local oscillator. sample volume determined injection beam waist was also investigated.

参考文章(2)
D. P. Hutchinson, K. L. Vander Sluis, J. Sheffield, D. J. Sigmar, Feasibility of alpha‐particle measurement by CO2 laser Thomson scattering Review of Scientific Instruments. ,vol. 56, pp. 1075- 1077 ,(1985) , 10.1063/1.1138512
S. Lee, T. Kondoh, Y. Yonemoto, Y. Miura, A filter bank system for scattered spectrum analysis in collective Thomson scattering diagnostic on JT-60 Review of Scientific Instruments. ,vol. 71, pp. 4445- 4448 ,(2000) , 10.1063/1.1326932