An Improved Time-Amplitude Window Discriminator

作者: M. J. Bak , E. M. Schmidt

DOI: 10.1109/TBME.1977.326198

关键词:

摘要: Usually extracellular microelectrodes record the activity from several neurons. In order to quantify of an individual neuron in some form spike discrimination is required. This paper describes a window discriminator for separation multi-unit neuronal trains which will display simultaneously recorded information and time-amplitude levels. The levels are displayed can be delayed include any portion waveform either above or below initial threshold level. Several instruments cascaded multiple interrogations same class single electrode recording. instrument utilizes inexpensive commercially available integrated circuits.

参考文章(6)
Edward M. Schmidt, R. Gilbert Jost, Kathleen K. Davis, Reexamination of the force relationship of cortical cell discharge patterns with conditioned wrist movements Brain Research. ,vol. 83, pp. 213- 223 ,(1975) , 10.1016/0006-8993(75)90931-2
R.G Dyer, P Beechey, A window discriminator, with visual display, suitable for unit recording Electroencephalography and Clinical Neurophysiology. ,vol. 31, pp. 621- 624 ,(1971) , 10.1016/0013-4694(71)90080-0
E.M. Schmidt, M.J. Bak, J.S. McIntosh, Long-term chronic recording from cortical neurons Experimental Neurology. ,vol. 52, pp. 496- 506 ,(1976) , 10.1016/0014-4886(76)90220-X
R. E. Daigle, R. J. Morgan, A. Budak, Window Discrimination Using Integrated Circuits IEEE Transactions on Biomedical Engineering. ,vol. BME-20, pp. 303- 306 ,(1973) , 10.1109/TBME.1973.324200
Edward M. Schmidt, An Instrument for Separation of Multiple-Unit Neuroelectric Signals IEEE Transactions on Biomedical Engineering. ,vol. BME-18, pp. 155- 157 ,(1971) , 10.1109/TBME.1971.4502816
R B Northrop, H J Grossman, An integrated-circuit pulse-height discriminator with multiplexed display. Journal of Applied Physiology. ,vol. 37, pp. 946- 950 ,(1974) , 10.1152/JAPPL.1974.37.6.946