An in vivo determination of total hip arthroplasty pistoning during activity.

作者: Adolph V. Lombardi , Thomas H. Mallory , Douglas A. Dennis , Richard D. Komistek , Robert A. Fada

DOI: 10.1054/ARTH.2000.6637

关键词:

摘要: The purpose of this study was to determine what extent hip joint separation occurs during normal gait on a treadmill and an abduction/adduction leg-lift maneuver in patients who have undergone total arthroplasty (THA). Eight had 10 successful unconstrained THAs (Harris Hip Scores >90) performed successive motions electronically powered leg lift while under fluoroscopy. fluoroscopic video images were analyzed using 3-dimensional model-fitting technique that converts 2-dimensional into real-time images. determined be present if the amount >0.75 mm, calculated linear error. During both activities, all experienced femoral head/acetabular component separation. For gait, maximum 2.8 minimum 0.8 mm (average, 1.2 mm). lift, 3.0 minimal 1.7 2.4 It appears head separates from acetabular but remains contact with superior-most tip component. Potential detrimental effects resulting include premature polyethylene wear loosening. Wear may enhanced because creation multidirectional vectors or excessive loads eccentric pivoting. These data valuable simulation studies better duplicate patterns observed retrieval analysis.

参考文章(45)
Georg Bergmann, Friedmar Graichen, Antonius Rohlmann, Holger Linke, Hip joint forces during load carrying Clinical Orthopaedics and Related Research. ,vol. 335, pp. 190- 201 ,(1997) , 10.1097/00003086-199702000-00018
J.P. PAUL, 31 – BIO-ENGINEERING STUDIES OF THE FORCES TRANSMITTED BY JOINTS: (II) Engineering Analysis Biomechanics and Related Bio-Engineering Topics#R##N#Proceedings of a Symposium Held in Glasgow, September 1964. pp. 369- 380 ,(1965) , 10.1016/B978-1-4831-6701-5.50040-1
Allan E. Gross, William N. Dust, Acute polyethylene fracture in an uncemented acetabular cup. Canadian Journal of Surgery. ,vol. 40, pp. 310- 312 ,(1997)
HA McKellop, IC Clarke, Degradation and Wear of Ultra-High-Molecular-Weight Polyethylene ASTM International. ,(1985) , 10.1520/STP33263S
Thomas Ilchmann, Radiographic assessment of cup migration and wear after hip replacement Acta Orthopaedica Scandinavica. ,vol. 276, pp. 1- 26 ,(1997) , 10.1080/17453674.1997.11744768
D T Davy, G M Kotzar, R H Brown, K G Heiple, V M Goldberg, K G Heiple, J Berilla, A H Burstein, Telemetric force measurements across the hip after total arthroplasty. Journal of Bone and Joint Surgery, American Volume. ,vol. 70, pp. 45- 50 ,(1988) , 10.2106/00004623-198870010-00008
MOTOI YAMAGUCHI, THOMAS W. BAUER, YASUSHI HASHIMOTO, Three-dimensional analysis of multiple wear vectors in retrieved acetabular cups. Journal of Bone and Joint Surgery, American Volume. ,vol. 79, pp. 1539- 1544 ,(1997) , 10.2106/00004623-199710000-00011
M. E. TORCHIA, R. A. KLASSEN, A. J. BIANCO, Total Hip Arthroplasty with Cement in Patients Less Than Twenty Years Old. Long-Term Results* Journal of Bone and Joint Surgery, American Volume. ,vol. 78, pp. 995- 1003 ,(1996) , 10.2106/00004623-199607000-00003
Thomas H. Mallory, Adolph V Lombardi, Robert A. Fada, Stephen M. Herrington, Robert W. Eberle, Dislocation after total hip arthroplasty using the anterolateral abductor split approach. Clinical Orthopaedics and Related Research. ,vol. 358, pp. 166- 172 ,(1999) , 10.1097/00003086-199901000-00020