Design of a system for contact-free measurement of the conductivity of biological tissue.

作者: M. Saint-George , C. H. Riedel , O. Dössel

DOI: 10.1515/BMTE.2002.47.S1B.794

关键词:

摘要: The electrical impedance of tissue can give important informations about the viability tissue. A non-invasive and contact-free measuring system using an inductive sensor is presented. Using a single pick-up coil not sensitive enough for small changes in conductivity biological Employing gradiometer instead improve resolution system. developed data acquisition realized four parts: lock-in-amplifier, providing sinusoidal signal sensor, power amplifier, driving excitation preamplifier buffering reference amplifying output sensor. In this paper focus on hardware that was set up. Results measurements inhomogeneous phantoms are shown.

参考文章(3)
Hermann Scharfetter, Helmut K Lackner, Javier Rosell, Magnetic induction tomography: hardware for multi-frequency measurements in biological tissues. Physiological Measurement. ,vol. 22, pp. 131- 146 ,(2001) , 10.1088/0967-3334/22/1/317
H Griffiths, Magnetic induction tomography Measurement Science and Technology. ,vol. 12, pp. 1126- 1131 ,(2001) , 10.1088/0957-0233/12/8/319