作者: Norimasa Yoza , Kouji Kouchiyama , Tohru Miyajima , Shiberu Ohashi
DOI: 10.1080/00032717508059036
关键词:
摘要: Abstract An atomic absorption flow detector combined with a gel chromatographic column (Sephadex G-25) gives sensitive and quantitative method of determining various condensed phosphates such as diphosphate, tri-phosphate, tetraphosphate Kurrol's salt. This is based on the automatic recording at resonance line magnesium due to complexes which are produced during elution phosphate anions through pre-equilibrated chloride solution.