Radiation dose to the patient in cardiac radiology

作者: G. M. Ardran , J. Hamill , E. Emrys-Roberts , R. Oliver

DOI: 10.1259/0007-1285-43-510-391

关键词:

摘要: Abstract In 54 patients the exposure at skin during fluoroscopy for cardiac catheterisation was found to average 21·4 R. The corresponding total area in R.cm2 averaged 1560 R.cm2. effective of irradiation calculated from these readings 117 cm2. results have been compared with a recent study by Gough et al. (1968), and implications irradiated are discussed relation bone marrow dose.