Absolute isotopic abundance ratios of three common lead reference samples

作者: E.J. Catanzaro

DOI: 10.1016/0012-821X(67)90057-X

关键词:

摘要: Abstract The absolute isotopic abundance ratios have been determined for three common lead reference samples, using solid-sample mass spectrometry. Samples of known composition, prepared from nearly pure separated 206Pb and 208Pb isotopes, were used to calibrate the spectrometers. resulting values are as follows: GS 4: 204Pb/206Pb = 0.061888 ± 0.000065, 207Pb/206Pb 0.95342 0.00036, 208Pb/206Pb 2.2181 0.0010; NBS 200: 0.068353 0.000072, 1.00538 0.00038, 2.3675 0.0011; C.I.T. Pb Std.: 0.060151 0.000063, 0.93081 0.00035, 2.1835 0.0010. indicated uncertainties overall limits error based on 95% confidence mean allowances effects sources possible systematic error.

参考文章(2)
E. J. Catanzaro, Triple-filament method for solid-sample lead isotope analysis Journal of Geophysical Research. ,vol. 72, pp. 1325- 1327 ,(1967) , 10.1029/JZ072I004P01325
E.J. Catanzaro, T.J. Murphy, W.R. Shields, E.L. Garner, Absolute isotopic abundance ratios of common, equal-atom, and radiogenic lead isotopic standards Journal of Research of the National Bureau of Standards Section A: Physics and Chemistry. ,vol. 72A, pp. 261- 267 ,(1968) , 10.6028/JRES.072A.025