An implantable optical blood pressure sensor based on pulse transit time

作者: Jens Fiala , Philipp Bingger , Dominic Ruh , Katharina Foerster , Claudia Heilmann

DOI: 10.1007/S10544-012-9689-9

关键词:

摘要: An implantable sensor system for long-term monitoring of blood pressure is realized by taking advantage the correlation between pulse transit time and pressure. The highly integrated module, fabricated using MEMS technologies, uses 8 light emitting diodes (LEDs) a photodetector on chip level. applied to large vessels, such as carotid or femoral arteries, allows extravascular measurement highly-resolved photoplethysmograms. In addition, spectrophotometric approaches allow hemoglobin derivatives. For calibration measurements, has been successfully implemented in animal models.

参考文章(38)
R. R. Coifman, D. L. Donoho, Translation-Invariant De-Noising Wavelets and Statistics. pp. 125- 150 ,(1995) , 10.1007/978-1-4612-2544-7_9
Michela Masè, Walter Mattei, Roberta Cucino, Luca Faes, Giandomenico Nollo, Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure. Journal of Electrocardiology. ,vol. 44, pp. 201- 207 ,(2011) , 10.1016/J.JELECTROCARD.2010.11.019
Joseph A. Potkay, Long term, implantable blood pressure monitoring systems Biomedical Microdevices. ,vol. 10, pp. 379- 392 ,(2008) , 10.1007/S10544-007-9146-3
Brian Gribbin, Andrew Steptoe, Peter Sleight, Pulse Wave Velocity as a Measure of Blood Pressure Change Psychophysiology. ,vol. 13, pp. 86- 90 ,(1976) , 10.1111/J.1469-8986.1976.TB03344.X
Andrew Steptoe, Harold Smulyan, Brian Gribbin, Pulse Wave Velocity and Blood Pressure Change: Calibration and Applications Psychophysiology. ,vol. 13, pp. 488- 493 ,(1976) , 10.1111/J.1469-8986.1976.TB00866.X
Paul A. Obrist, Kathleen C. Light, James A. McCubbin, J. Stanford Hutcheson, J. Lee Hoffer, Pulse Transit Time: Relationship to Blood Pressure and Myocardial Performance Psychophysiology. ,vol. 16, pp. 292- 301 ,(1979) , 10.1111/J.1469-8986.1979.TB02993.X
J. Martin Bland, DouglasG. Altman, Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. ,vol. 327, pp. 307- 310 ,(1986) , 10.1016/S0140-6736(86)90837-8
Mico Yee-Man Wong, Carmen Chung-Yan Poon, Yuan-Ting Zhang, An Evaluation of the Cuffless Blood Pressure Estimation Based on Pulse Transit Time Technique: a Half Year Study on Normotensive Subjects Cardiovascular Engineering. ,vol. 9, pp. 32- 38 ,(2009) , 10.1007/S10558-009-9070-7
Anders Johansson, C Ahlstrom, T Lanne, Per Ask, None, Pulse wave transit time for monitoring respiration rate. Medical & Biological Engineering & Computing. ,vol. 44, pp. 471- 478 ,(2006) , 10.1007/S11517-006-0064-Y