Adaptive Cancellation of Floor Vibrations in Standing Ballistocardiogram Measurements Using a Seismic Sensor as a Noise Reference

作者: O.T. Inan , M. Etemadi , B. Widrow , G.T.A. Kovacs

DOI: 10.1109/TBME.2009.2018831

关键词:

摘要: An adaptive noise canceller was used to reduce the effect of floor vibrations on ballistocardiogram (BCG) measurements from a modified electronic bathroom scale. A seismic sensor placed next scale and as reference input canceller. BCG recordings were acquired healthy subject while another person stomped around scale, thus causing increased vibrations. The substantially eliminated artifacts in signal due these without distorting morphology measured BCG. Additionally, obtained standing inside parked bus engine running. engine, other vehicles moving road bus, also effectively by system with automatic vibration cancellation could be increase measurement robustness home monitoring applications. approach may enable recording ambulances-or transport vehicles-where noninvasive hemodynamic otherwise not feasible.

参考文章(12)
Jim Williams, Bridge circuits Analog Circuit Design. pp. 638- 678 ,(2011) , 10.1016/B978-0-12-385185-7.00032-9
William Dock, Felix Taubman, Some technics for recording the ballistocardiogram directly from the body The American Journal of Medicine. ,vol. 7, pp. 751- 755 ,(1949) , 10.1016/0002-9343(49)90413-1
O T Inan, M Etemadi, R M Wiard, L Giovangrandi, G T A Kovacs, Robust ballistocardiogram acquisition for home monitoring. Physiological Measurement. ,vol. 30, pp. 169- 185 ,(2009) , 10.1088/0967-3334/30/2/005
A. J. Williams, G. C. Mergner, A Self-Balancing Bridge with Range and Precision Adequate for a Platinum Thermometer IEEE Transactions on Instrumentation and Measurement. ,vol. 15, pp. 121- 131 ,(1966) , 10.1109/TIM.1966.4313521
B. Widrow, J.R. Glover, J.M. McCool, J. Kaunitz, C.S. Williams, R.H. Hearn, J.R. Zeidler, Jr. Eugene Dong, R.C. Goodlin, Adaptive noise cancelling: Principles and applications Proceedings of the IEEE. ,vol. 63, pp. 1692- 1716 ,(1975) , 10.1109/PROC.1975.10036
Omer T. Inan, Mozziyar Etemadi, Richard M. Wiard, Gregory T. A. Kovacs, Laurent Giovangrandi, Non-invasive measurement of Valsalva-induced hemodynamic changes on a bathroom scale Ballistocardiograph international conference of the ieee engineering in medicine and biology society. ,vol. 2008, pp. 674- 677 ,(2008) , 10.1109/IEMBS.2008.4649242
Omer T. Inan, Mozziyar Etemadi, Richard M. Wiard, Laurent Giovangrandi, Gregory T. A. Kovacs, Evaluating the Foot Electromyogram Signal as a Noise Reference for a Bathroom Scale Ballistocardiogram Recorder computer-based medical systems. pp. 70- 74 ,(2008) , 10.1109/CBMS.2008.78
Isaac Starr, A. J. Rawson, H. A. Schroeder, N. R. Joseph, STUDIES ON THE ESTIMATION OF CARDIAC OUPTUT IN MAN, AND OF ABNORMALITIES IN CARDIAC FUNCTION, FROM THE HEART'S RECOIL AND THE BLOOD'S IMPACTS; THE BALLISTOCARDIOGRAM American Journal of Physiology-Legacy Content. ,vol. 127, pp. 1- 28 ,(1939) , 10.1152/AJPLEGACY.1939.127.1.1
J. Alihanka, K. Vaahtoranta, I. Saarikivi, A new method for long-term monitoring of the ballistocardiogram, heart rate, and respiration American Journal of Physiology-regulatory Integrative and Comparative Physiology. ,vol. 240, ,(1981) , 10.1152/AJPREGU.1981.240.5.R384
T. Koivistoinen, S. Junnila, A. Varri, T. Koobi, A new method for measuring the ballistocardiogram using EMFi sensors in a normal chair international conference of the ieee engineering in medicine and biology society. ,vol. 3, pp. 2026- 2029 ,(2004) , 10.1109/IEMBS.2004.1403596