Experimental and Numerical Analyses of the Pull-out Response of a Steel Post/Bovine Bone Cementless Fixation

作者: Khaled Gammoudi , Mohamed Kharrat , Maher Dammak

DOI: 10.1016/S1672-6529(11)60142-X

关键词:

摘要: Effect of initial interference fit on pull-out strength in cementless fixation between bovine tibia and smooth stainless steel post was investigated this study. Compressive behavior spongious bone studied using mechanical testing order to evaluate the elastic-plastic properties different regions proximal tibia. Friction tests were carried out aim friction contact steel. A cylindrical inserted a pre-drilled with an taken as vitro model assess contribution stability Total Knee Arthroplasty (TKA) tibial component. Pull-out experiments for fits. Finite Element Models (FEM) local developed analysis experimental ultimate force results. At post/bone interface, Coulomb considered FEM calculations pressure-dependent coefficient. It found that results are good agreement The interfacial stresses indicates micro-slip initiation depends properties.

参考文章(22)
K. Gammoudi, M. Kharrat, M. Dammak, R. Abdelmoula, S. Ramtani, Pull-Out Response of a Steel Post Inserted in a Pre-Drilled HDPE Cylinder: Analytical and Finite Element Analyses Using Pressure-Dependent Friction Journal of Adhesion Science and Technology. ,vol. 26, pp. 1157- 1167 ,(2012) , 10.1163/156856111X593559
P. Lasaygues, M. Pithioux, Ultrasonic characterization of orthotropic elastic bovine bones. Ultrasonics. ,vol. 39, pp. 567- 573 ,(2002) , 10.1016/S0041-624X(02)00261-5
Aaron A. Hofmann, Tyler D. Goldberg, Amie M. Tanner, Thomas M. Cook, Surface Cementation of Stemmed Tibial Components in Primary Total Knee Arthroplasty The Journal of Arthroplasty. ,vol. 21, pp. 353- 357 ,(2006) , 10.1016/J.ARTH.2005.06.012
Dan-Jae Lin, Chien-Ping Ju, Shu-Huei Huang, Yin-Chun Tien, Hsiang-Shu Yin, Wen-Cheng Chen, Jiin-Huey Chern Lin, Mechanical testing and osteointegration of titanium implant with calcium phosphate bone cement and autograft alternatives. Journal of The Mechanical Behavior of Biomedical Materials. ,vol. 4, pp. 1186- 1195 ,(2011) , 10.1016/J.JMBBM.2011.04.001
Ian P Holloway, Patrick J Lusty, William L Walter, William K Walter, Bernie A Zicat, None, Tibial Fixation Without Screws in Cementless Knee Arthroplasty The Journal of Arthroplasty. ,vol. 25, pp. 46- 51 ,(2010) , 10.1016/J.ARTH.2008.09.023
Frank Linde, Ivan Hvid, Frank Madsen, The effect of specimen geometry on the mechanical behaviour of trabecular bone specimens. Journal of Biomechanics. ,vol. 25, pp. 359- 368 ,(1992) , 10.1016/0021-9290(92)90255-Y
Mariya Poukalova, Christopher M. Yakacki, Robert E. Guldberg, Angela Lin, Minn Saing, Scott D Gillogly, Ken Gall, Pullout strength of suture anchors: effect of mechanical properties of trabecular bone. Journal of Biomechanics. ,vol. 43, pp. 1138- 1145 ,(2010) , 10.1016/J.JBIOMECH.2009.12.007
Tony M. Keaveny, Robert E. Borchers, Lorna J Gibson, Wilson C. Hayes, Trabecular bone modulus and strength can depend on specimen geometry. Journal of Biomechanics. ,vol. 26, pp. 991- 1000 ,(1993) , 10.1016/0021-9290(93)90059-N