Multi-loop relaxation oscillation SQUID magnetometers with large flux-to-voltage transfer functions

作者: J. Kawai , G. Uehara , M. Higuchi , H. Ogata , H. Kado

DOI: 10.1109/77.919576

关键词:

摘要: A multi-loop Relaxation Oscillation SQUID (ROS) with direct voltage readout is fabricated and characterized. We designed four-loop ROSs the size of 5 mm /spl times/5 mm. maximum modulation 100 mu/V a flux-to-voltage transfer function more than 10 mV//spl Phi//sub 0/ were obtained. With flux locked loop (FLL), magnetic field resolution 4 to fT//spl radic/Hz was obtained in frequency range from 0.3 Hz up white region. Using this magnetometer, we demonstrated measurement brain response.

参考文章(7)
Derk Jan Adelerhof, Jun Kawai, Keiji Tsukada, Gen Uehara, Hisashi Kado, Magnetometers based on double relaxation oscillation superconducting quantum interference devices Applied Physics Letters. ,vol. 66, pp. 2274- 2276 ,(1995) , 10.1063/1.113190
Gen Uehara, Naoki Matsuda, Kunio Kazami, Youichi Takada, Hisashi Kado, Asymmetric Bias Injection Technique for Drung-Type Superconducting Quantum Interference Devices Japanese Journal of Applied Physics. ,vol. 32, pp. L1735- L1738 ,(1993) , 10.1143/JJAP.32.L1735
D. Drung, S. Knappe, H. Koch, THEORY FOR THE MULTILOOP DC SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE MAGNETOMETER AND EXPERIMENTAL VERIFICATION Journal of Applied Physics. ,vol. 77, pp. 4088- 4098 ,(1995) , 10.1063/1.359607
Derk Jan Adelerhof, Jun Kawai, Gen Uehara, Hisashi Kado, High sensitivity double relaxation oscillation superconducting quantum interference devices with large transfer from flux to voltage Review of Scientific Instruments. ,vol. 66, pp. 2631- 2637 ,(1995) , 10.1063/1.1145600
H. Kado, M. Higuchi, M. Shimogawara, Y. Haruta, Y. Adachi, J. Kawai, H. Ogata, G. Uehara, Magnetoencephalogram systems developed at KIT IEEE Transactions on Applied Superconductivity. ,vol. 9, pp. 4057- 4062 ,(1999) , 10.1109/77.783918
D.J. Adelerhof, H. Nijstad, J. Flokstra, H. Rogalla, Relaxation oscillation SQUIDs with high delta V/ delta Phi IEEE Transactions on Applied Superconductivity. ,vol. 3, pp. 1862- 1865 ,(1993) , 10.1109/77.233318
J.P. Wikswo, SQUID magnetometers for biomagnetism and nondestructive testing: important questions and initial answers IEEE Transactions on Applied Superconductivity. ,vol. 5, pp. 74- 120 ,(1995) , 10.1109/77.402511