MWCNT-cysteamine-Nafion modified gold electrode based on myoglobin for determination of hydrogen peroxide and nitrite.

作者: Erhan Canbay , Berika Şahin , Müge Kıran , Erol Akyilmaz

DOI: 10.1016/J.BIOELECHEM.2014.09.001

关键词:

摘要: Abstract In this work, a novel amperometric biosensor of hydrogen peroxide (H 2 O ) was developed based on the immobilization myoglobin (Mb) surface multi-walled carbon nanotube (MWCNT) –Nafion–cysteamine (CA) modified gold electrode (Au) and its electrocatalytic activity used for determination nitrite (NO − ). optimization studies, best MWCNT amount were investigated. It discovered at experiments working conditions that buffer study as 50.0 mM, pH 7.0 phosphate (PBS) temperature 30 °C H biosensor. determined characterization studies linear results are obtained between ranges 0.1 μM to 70.0 μM concentration 1–250 μM NO . The reproducibility both nitrite. From experiments, average value, standard deviation (SD) coefficients variation (CV%) calculated be 10.02 ± 0.43 μM, 4.29% 10.0 μM ( n  = 6) 52.0 ± 2.1 μM, 3.89% 50.0 μM  = 8), respectively. At same time sample analyzed in drinking mineral waters.

参考文章(33)
C.P. Andrieux, P. Audebert, B. Divisia-Blohorn, P. Aldebert, F. Michalak, Electrochemistry in hydrophobic Nafion gels Journal of Electroanalytical Chemistry and Interfacial Electrochemistry. ,vol. 296, pp. 117- 128 ,(1990) , 10.1016/0022-0728(90)87237-E
Craig E. Banks, Ryan R. Moore, Trevor J. Davies, Richard G. Compton, Investigation of modified basal plane pyrolytic graphite electrodes: definitive evidence for the electrocatalytic properties of the ends of carbon nanotubes. Chemical Communications. pp. 1804- 1805 ,(2004) , 10.1039/B406174H
Ariel Garcia, Carlos Peniche-Covas, Belkis Chico, Benjamin K. Simpson, Reynaldo Villalonga, Ferrocene branched chitosan for the construction of a reagentless amperometric hydrogen peroxide biosensor. Macromolecular Bioscience. ,vol. 7, pp. 435- 439 ,(2007) , 10.1002/MABI.200700032
Sheying Dong, Nan Li, Tinglin Huang, Hongsheng Tang, Jianbin Zheng, Myoglobin immobilized on LaF3 doped CeO2 and ionic liquid composite film for nitrite biosensor Sensors and Actuators B-chemical. ,vol. 173, pp. 704- 709 ,(2012) , 10.1016/J.SNB.2012.07.073
WILLIAM LIJINSKY, SAMUEL S. EPSTEIN, Nitrosamines as environmental carcinogens. Nature. ,vol. 225, pp. 21- 23 ,(1970) , 10.1038/225021A0
V. V. Kuznetsov, S. V. Zemyatova, Flow-injection spectrophotometry of nitrites based on the diazotization reactions of azine dyes Journal of Analytical Chemistry. ,vol. 62, pp. 637- 644 ,(2007) , 10.1134/S1061934807070052
Xiao-Fei Wang, Jing-Juan Xu, Hong-Yuan Chen, A New Electrochemiluminescence Emission of Mn2+-Doped ZnS Nanocrystals in Aqueous Solution Journal of Physical Chemistry C. ,vol. 112, pp. 17581- 17585 ,(2008) , 10.1021/JP807136S